62
Amplitude
Chirp
Template
Constant
Set to Hold
Tip:
INTEGRATED PATCH AMPLIFIER OPERATION - M
Pulse Width (ms)
Segment Duration
[Same as the Waveform Editor / Duration field]
Live Preview
This waveform generates a sinusoidal wave that
increases in frequency over time.
Chirp Type
Amplitude (mV)
Start Frequency (Hz) 1 – 50,000
End Frequency (Hz)
Segment Duration
[Same as the Waveform Editor / Duration field]
Live Preview
Select the waveform amplitude for a Segment
Use the IPA Control panel holding level for the
Segment's amplitude.
When used in Segment 1 for voltage-clamp experiments,
this allows you to record the leak current along with the
actual holding voltage, For current-clamp experiments,
this allows you to record the actual cell potential and
holding current.
When used in the last Segment, information after
stimulation is recorded, such as for tail currents.
– R
ANUAL
Duration of second pulse
Square wave duration
Update the Routine Editor
preview in real time.
Linear
A linear increase in
frequency
Geometric
A geometric increase
in frequency
±1000
1 – 50,000
Chirp wave duration
Update the Routine Editor
preview in real time.
. 1.0.6 (20160729)