Describing The Control Word - SCHNIER smart-E 310 Operating Manual

High-voltage supply
Table of Contents

Advertisement

11.2.5.
Byte
Bit
0
0
High voltage ON
1
Reset error
2
Reserve
3
Reserve
4
Heartbeat
5
Reserve
6
Reserve
7
Relay test bit
1
0...7 Reserve
Operating manual 810378-BAL-EN-170330

Describing the control word

Control bit
Mach/30 2017
Description
1: Switch on high voltage
0: Switch off high voltage
1: All errors are confirmed
0: No action
Note: In order to switch on HV
again, after an error has been
confirmed, the "HV ON" bit must be
reset/set.
The HS-supply expects a heartbeat
signal (1 Hz) via EtherCAT-
controller.
This monitoring is initially deactivat-
ed and is then activated with the first
heartbeat.
If a heartbeat never comes then the
generator can be operated without it.
As soon as the heartbeat comes
once, this is monitored and if it
remains off the generator shows an
error.
1: Open the safety relay
0: No action / close relay
Note: The safety relay is controlled
by the generator depending on the
remaining voltage on the HV
connection.
If the contact is found in the "safe"
position (closed), it can be opened
with this Bit for test purposes.
Note: When the relay test bit is set,
the HV cannot be switched on.
Page 50 of 54