Describing The Control Word - SCHNIER smart-E 310 Operating Manual

High-voltage supply
Table of Contents

Advertisement

Example: From a smart-E 510 received data:
Control word
HV ON
Byte 0
Byte 1
0x01
0x00
kV ramp
10 kV/s
Byte
Byte
10
11
0x0a
0x00
Example: From a smart-E 510 sent data:
Status word
Byte 0
Byte 1
0xC1
0x00
10.2.4.
Byte
Bit
0
0
High voltage ON
1
Reset error
2
Reserve
3
Reserve
4
Heartbeat
5
Reserve
6
Reserve
7
Relay test bit
1
0...7 Reserve
Operating manual 810378-BAL-EN-170330
U Target value
U min
20 kV
5 kV
Byte 2
Byte 3
Byte 4
0x14
0x00
0x05
Sensitivity*
5
Byte
Byte
12
13
0x05
0x00
Service code
Byte 2
Byte 3
Byte 4
0x00
0x00
0x00

Describing the control word

Control bit
Description
1: Switch on high voltage
0: Switch off high voltage
1: All errors are confirmed
0: No action
Note: In order to switch on HV again, after an
error has been confirmed, the "HV ON" bit must
be reset/set.
The smart-E 310 or smart-E 510 expects a heart-
beat signal (1Hz) via EIP-controller.
This monitoring is initially deactivated and is then
activated with the first heartbeat.
If a heartbeat never comes, then the generator
can be operated without it.
As soon as the heartbeat comes once, this is
monitored and if it remains off the generator
shows an error.
1: Open the safety relay
0: No action / close relay
Note: The safety relay is controlled by the gener-
ator depending on the remaining voltage on the
HV connection.
If the contact is found in the "safe" position
(closed), it can be opened with this Bit for test
purposes.
Note: When the relay test bit is set, the HV can-
not be switched on.
I Limit
256 uA
Byte 5
Byte 6
Byte 7
0x00
0x00
0x01
U actual
20 kV
Byte 5
Byte 6
Byte 7
0x00
0xC8
0x00
Mach/30 2017
Imax
50 uA
Byte 8
Byte 9
0x32
0x00
I actual
10 uA
Byte 8
Byte 9
0x0A
0x00
Page 45 of 54