Specification of the Ethernet/IP interface

(smart-E 310e, smart-E510e, smart-E 310De, smart-E 510De)

10.1.Bus interface

Ethernet/IP based on the ethernet physical layer. To create communication between
the smart-E510 (EIP adaptor) and the Ethernet/IP controller (EIP scanner), the corre-
sponding settings must be adjusted in the controller (see manual of your Ethernet/IP
controller for further information):
The HV generator has a factory preset IP address 192.168.100.122. This can be
changed via the integrated webserver at http://IP-Adresse/ipconfig.
CAUTION
Changes should only be made here if you are knowledgeable in working
with network settings. If the settings no (longer) correspond to your net-
work and the connection can no longer be made, the device must be
returned to the manufacturer!
This page requires a username and password: (User/User).
For older devices, either no log in is needed or (admin/admin).
The HV generator is set permanently to 100 Mbit/s full duplex. The receiv-
er should also be permanently set to full duplex, as the "auto-negotiation"
setting responds to most bus participants to permanently full duplex with
"half duplex". This leads to sporadically repeating communication malfunc-
tions (incl. after several hours)
Operating manual 810378-BAL-EN-170330