Receive I/O Data Mapping (Pn-Io Controller To Pn-Io Device); I/O Data Mapping Transfer: (Pn-Io Device To Pn-Io Controller); I/O Data Definition (Area, Unit, Standard) - SCHNIER smart-E 310 Operating Manual

High-voltage supply
Table of Contents

Advertisement

9.2.3.
Byte 0
Byte 1
Control word
Byte 10~19
Reserve
9.2.4.
Byte 0
Byte 1
Status word
Byte 10~15
Reserve
9.2.5.
Data
Unit
U target
kV
U min
kV
I Limit*
uA
I max.*
uA
U actual
100
V
I actual
uA
*: Valid from SW version 3.0, prior to that: 550 uA
Values outside of these areas (min, max) are ignored. The warning "invalid parame-
ters" is output.
Caution: The Byte sequence is Little-Endian. The Byte sequence can be taken from the
following examples:
Example: From a smart-E 510 received data:
Control word
HV ON
Byte 0
Byte 1
0x01
0x00
Example: From a smart-E 510 sent data:
Byte 0
Byte 1
0xC0
0x00
Status word
Operating manual 810378-BAL-EN-170330

Receive I/O data mapping (PN-IO controller to PN-IO device)

Byte 2
Byte 3
Byte 4
U target value
U min (kV)
(kV)

I/O data mapping transfer: (PN-IO device to PN-IO controller)

Byte 2
Byte 3
Byte 4
Service code
Byte 16
SW Subversion
SW Version

I/O data definition (area, unit, standard)

Minimum
Maximum
10
100
10
100
10
800 / 600
10
800
0
1240
0
1130
U Target value
U min
20 kV
5 kV
Byte 2
Byte 3
Byte 4
0x14
0x00
0x05
Byte 2
Byte 3
Byte 4
0x00
0x00
0x00
Service code
Byte 5
Byte 6
I Limit (uA)
Byte 5
Byte 6
U actual (kV)
Byte 17
Byte 18
HW Version
Stand-
Remarks
ard
0
kV target value/limiting value
0
Limit value for Umin-switch off
0
Target value for current limit-
ing value (800 below 80 kV)
0
Limit value for Imax - switch
off
-
Actual value kV
-
Actual current
I Limit
256 uA
Byte 5
Byte 6
0x00
0x00
Byte 5
Byte 6
0x00
0xC8
U actual
20 kV
Mach/30 2017
Byte 7
Byte 8
Byte 9
Imax (uA)
Byte 7
Byte 8
Byte 9
I actual (uA)
Byte 19
Reserve
Imax
50 uA
Byte 7
Byte 8
Byte 9
0x01
0x32
0x00
Byte 7
Byte 8
Byte 9
00
0x0A
0x00
I actual
10 uA
Page 39 of 54