Electromagnetic compatibility (EMC)

Caution
 Uscan contains sensitive components and circuits. Failure to observe proper static control procedures
may result in damage to the system. Any faults should be reported to
distributor for repair.
Warning
 Uscan complies with the Electromagnetic Compatibility requirements of AS/NZ CISPR 11:2004 and
IEC 60601-1-2:2007. However, electronic and mobile communications equipment may transmit
electromagnetic energy through air and there is no guarantee that interference will not occur in a
particular installation or environment. Interference may result in artifacts, distortion, or degradation of
the ultrasound image. If Uscan is found to cause or respond to interference, try re-orienting Uscan or
the affected device, or increasing the separation distance between the devices. Contact
customer support or your Signostics distributor for further information.
The Uscan is intended for use in the electromagnetic environment specified below. The user of the Uscan
should assure that it is used in such an environment.
Table 8.1 Guidance and manufacturer's declaration: electromagnetic emissions
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
1
Uscan has Class A compliance in meaning it is suitable for use in all establishments other than domestic and those directly connected to the
public low-voltage power supply network that supplies buildings used for domestic purposes. If Uscan is found to cause or respond to
interference follow the guidelines in the warning section above
Table 8.2 Guidance and manufacturer's declaration: electromagnetic immunity
The user should assure the Uscan is used in the electromagnetic environment specified below.
Immunity test
IEC 60601 test level Compliance level
Electrostatic
±6 kV contact
discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
Surge
± 1kV line(s) to line(s)
IEC 61000-4-5
± 2kV line(s) to earth
Voltage dips, short
<5% U
interruptions and
U
) for 0.5 cycle
T
voltage variations on
40% U
power supply input
U
) for 5 cycles
T
lines
70% U
IEC 61000-4-11
for 25 cycles
<5% U
U
) for 5 sec
T
P003948
Electromagnetic environment: guidance
Uscan uses RF energy only for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
Uscan is suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
±6 kV contact ±8kV air Floors should be wood, concrete or ceramic tile. If floors
±2 kV for power supply
lines
± 1kV differential mode
± 2kV common mode
1
1
(>95% dip in
<5% U
(>95% dip in
T
T
U
) for 0.5 cycle
T
(60% dip in
40% U
(60% dip in
T
T
U
) for 5 cycles
T
(30% dip in U
70% U
(30% dip in U
T
T
T
for 25 cycles
(>95% dip in
<5% U
(>95% dip in
T
T
U
) for 5 sec
T
Uscan User Manual
Signostics
Electromagnetic environment: guidance
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
T
or your Uscan
Signostics
Page | 44