Battery Charger / Power Supply Operation; Operational Sequence - Accelerated Care Plus OMNISTIM FX2 PRO User Manual

Electrotherapy system
Table of Contents

Advertisement

12

Battery Charger / Power Supply Operation

Before using the Omnistim
Verify that rechargeable batteries are installed within the Omnistim
Check to see that the battery type selector switch located inside the battery compartment has been switched
to the rechargeable position.
If you plug in the charger, after two seconds, the charge LED on the Omnistim
batteries are fully charged, the LED blinks. If you plug in the charger during treatment or when the unit is ON the
LED turns off. (The charging continues but at a trickle and does not cause the LED to glow.) When you turn OFF
the unit after two seconds the LED comes on to announce charging.
NOTE: If non-rechargeable batteries are used:
1.
Do not use the charger.
2.
Make sure that the rechargeable/non-rechargeable selector switch in the bottom of the battery
compartment is in the non-rechargeable position.
Life expectancies of batteries under nominal (40mA MFAC) load conditions:
4.4Ah NiCad rechargeable batteries - 10 Hours.
Alkaline non-rechargeable batteries - 100 Hours.

Operational Sequence

1.
Press the POWER button (1/5) to turn on the power. The display panel will illuminate. There will be no
stimulation current and the unit will be in the SELECT PROTOCOL / TREATMENT OFF mode.
2.
At this time select the treatment protocol desired and the appropriate treatment program by using the PROTOCOL
buttons (2) to select PAIN, NEURO RE-ED, MUSCLE RE-ED, FUNCTIONAL RE-ED, or FUNCTIONAL RE-ED
ASSISTED.
3.
Press the SELECT PARAM button to modify any settings. Press the MODIFY TIME/PARAM UP of DOWN
buttons to adjust the desired treatment time, or other settings, if you wish to change from the default settings.
Set the timer for the desired treatment time. Pressing MODIFY TIME/PARAM UP or DOWN arrow buttons
increases or decreases the duration from a minimum of 1 minute to a maximum of 60 minutes. The timer will
count down to zero during treatment. When zero is reached, a warning tone is emitted from the unit, the
treatment current is switched off and TREATMENT OFF is displayed. Pressing the START/STOP button (4)
will shut off the warning tone. If the unit is left unattended, the warning tone is emitted for ten seconds before
power is automatically switched off. Note that the treatment time is pre-set for each program but may be
adjusted at will during treatment.
4.
Now position the electrodes on the patient, and connect the electrode plug(s). Adjust the output by pressing the
OUTPUT A and OUTPUT B UP or DOWN buttons (3), to deliver appropriate output based on patient
sensation. The screen will display the following (in this example, TENS-LVPC MOTOR was selected from the
NEURO RE-ED protocol functional group.
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
®
2
FX
Pro battery charger/power supply:
NEURO RE-ED
TENS-LVPC MOTOR
SET OUTPUTS A-B
A: 19 mA
B: 19 mA
®
FX
200 µS
®
2
OMNISTIM
FX
Pro USER MANUAL
2
Pro battery compartment.
®
2
FX
Pro comes on. After the