Factory Settings - Accelerated Care Plus OMNISTIM FX2 PRO User Manual

Electrotherapy system
Table of Contents

Advertisement

®
2
OMNISTIM
FX
PRO USER MANUAL
K. Output for CHANNEL A.
L. Output for CHANNEL B.
M. Increase / decrease OUTPUT A. Press UP arrow button to increase intensity in channel A. Press the DOWN
arrow button to decrease intensity in channel A.
N. Increase / decrease OUTPUT B. Press UP arrow button to increase intensity in channel B. Press the DOWN
arrow button to decrease intensity in channel B.
O. LED which lights-up red when output is being applied thru channel B.
P. LED which lights-up red when output is being applied thru channel A.
Q. START/STOP button. Having selected the type of treatment and protocol, press to start or stop the treatment.
R. LCD graphics screen for all functions and parameters. The graphics screen displays:
Protocol selected
Program selected
Time remaining
Carrier Frequency
Bar graph for Vector and detection modes
S. PROTOCOL LABEL to assist therapist in quick operation and treatment selection.
T. Power Supply input jack.
U. Power Supply operation indicator LED. This light is on when the power supply is plugged in and power is
supplied to the unit.
V. Solid carrying handle for easy transport of the unit.

Factory Settings

®
2
The Omnistim
FX
Pro comes with the following factory settings:
These are generally suitable for most clinicians. To adjust these default settings proceed as follows:
1.
When the main screen comes up and displays "SELECT PROTOCOL / TREATMENT OFF" push the SELECT
PARAMETER button. You will then be able to adjust the factory settings. The new settings will be saved for
use as long as you do not remove the batteries or allow a full discharge of the unit.
2.
The BUTTON SPEED for output A & B sets the speed at which output will increase and decrease when you
push the buttons. This should generally be set at a slower speed (85 - 95) so as not to startle the patient.
3.
AUDIO on or off disables or enables the audio system, which beeps at the end of the treatment and during the
NMES programs.
NOTE: The above settings cannot be accessed during operation and are only available following the start up
screen. The settings will reset to their initial default parameters if the batteries are removed or the unit is fully
discharged. To restore the new settings follow the procedure in Appendix I.
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Button Speed, A & B Output
Audio
Button speeds
Vector selection
Output current
Sweep rate and modulation selection
87
ON
11