Functional Re-Ed - Accelerated Care Plus OMNISTIM FX2 PRO User Manual

Electrotherapy system
Table of Contents

Advertisement

- 32 -
B. LVPC SURFACE MUSCLE
Program used for strength development and treatment of muscle disuse atrophy. Activation of muscle with
electrical stimulation at a high intensity for a short time with long "ON/OFF" ramps reduces atrophy and
improves strength.
stimulation "ON" time. Target tissues are smaller superficial muscle groups. Bipolar set-up over agonist
and antagonist muscles for target muscle. The intensity should be set to elicit a grade 2 to 4 muscle
contraction. Treatment time should be reduced based on muscle fatigue. The affected joint movement
should be restricted by holding manually, or using weights or exercise bands.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
Y
N
N
N
Y
2
Y
Y
Y
Y
Y
Y
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.

Functional Re-Ed

A. SLOW CYCLE
Program simulates slow cycling, at 42 rpm, and can be used to retrain reciprocal movement of the L.E.
with or without resistance. Electrodes are applied to both quads while sitting with hips and knees flexed.
Use a lower extremity ergometer or pedal exerciser in combination with stimulation at mild to moderate
contraction intensity for closed chain functional training. Increase resistance as patient progresses in the
rehab program.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
N
N
Y
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
The patient should participate to the extent possible by contracting during the
Parameter
Name
6
PHASE DURATION
7
PULSE RATE
PHASE DURATION MOD
MODULATION
RATE SCAN TIME
1
TREATMENT TIME
2
9
B OUTPUT DELAY
4
RAMP UP
5
RAMP DOWN
2
T-ON
3
T-OFF
8
CHANNEL MODE
Parameter
Name
2
PHASE DURATION
CYCLE TIME
PULSE RATE
1
TREATMENT TIME
OMNISTIM
Adjustable
Range (unit)
40...300 (μs)
0.0...250 (Hz)
0...99 (min)
0.0...9.9 (s)
0.0...9.9 (s)
0.0...9.9 (s)
0...30 (s)
0...199 (s)
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
Adjustable
Range (unit)
40...100 (μs)
0...99 (min)
®
2
FX
PRO USER MANUAL
Default (unit)
70 (μs)
50.0 (Hz)
0
OFF
0 (s)
10 (min)
0.0 (s)
2.0 (s)
2.0 (s)
1
10 (s)
1
50 (s)
A/B SIM.
Default (unit)
70 (μs)
1.40 (s)
50 (Hz)
15 (min)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

100fx2c

Table of Contents