Muscle Re-Ed - Accelerated Care Plus OMNISTIM FX2 PRO User Manual

Electrotherapy system
Table of Contents

Advertisement

®
2
OMNISTIM
FX
PRO USER MANUAL
E. MFAC RECIPROCAL INHIBITION
This program provides for stimulation of the spastic muscle's antagonists. It activates reciprocal inhibition
of the spastic muscle reducing tone. Slow ramps decrease the potential to trigger spasticity of the agonist.
A single channel (A or B) is used over the spastic muscle's antagonist. Intensity should be set to elicit a
grade 2 to 3 muscle contraction. Intensity should not be too high to induce overflow activation of the
spastic muscle. Treatment time can be reduced based on muscle fatigue and inhibition of spasticity.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
Y
N
N
N
Y
2
Y
Y
Y
Y
Y
Y
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.

Muscle Re-Ed

A. MFAC SURFACE DEEP MUSCLE
Program used for strength development and treatment of muscle disuse atrophy. Activation of muscle with
electrical stimulation at a high intensity for a short time with long "ON/OFF" ramps reduces atrophy and
improves strength. The patient should participate to the extent possible by contracting during the stimulation
"ON" time. Target tissues are superficial and deep muscle groups. Bipolar set-up over agonist and antagonist
muscles for target muscle. The intensity should be set to elicit a grade 2 to 4 muscle contraction. Treatment
time should be reduced based on muscle fatigue. The affected joint movement should be restricted by holding
manually, or using weights or exercise bands.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
Y
N
N
N
Y
2
Y
Y
Y
Y
Y
Y
1
When T-OFF = 0, T-ON = CONTINUOUS.
2
B OUTPUT DELAY parameter is shown only when the Channel Mode is set to B DELAYED.
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Parameter
Name
2
CARRIER FREQ
3
BURST FREQUENCY
BURST RATE SCAN
RATE SCAN
RATE SCAN TIME
1
TREATMENT TIME
2
9
B OUTPUT DELAY
6
RAMP UP
7
RAMP DOWN
4
T-ON
5
T-OFF
8
CHANNEL MODE
Parameter
Name
9
CARRIER FREQ
6
BURST FREQUENCY
BURST RATE SCAN
RATE SCAN
RATE SCAN TIME
1
TREATMENT TIME
2
8
B OUTPUT DELAY
4
RAMP UP
5
RAMP DOWN
2
T-ON
3
T-OFF
7
CHANNEL MODE
Adjustable
Range (unit)
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
0.0...250 (Hz)
0...99 (min)
0.0...9.9 (s)
0.0...9.9 (s)
0.0...9.9 (s)
0...30 (s)
0...199 (s)
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
Adjustable
Range (unit)
2.0, 2.5, 4.0, 5.0, 10.0 (kHz)
0.0...250 (Hz)
0...99 (min)
0.0...9.9 (s)
0.0...9.9 (s)
0.0...9.9 (s)
0...30 (s)
0...199 (s)
A/B SIMULTANEOUS, A/B ALTERNATE, B DELAYED
- 31 -
Default (unit)
2.5 (kHz)
50 (Hz)
20 (%)
ON
4 (s)
15 (min)
0.0 (s)
6.0 (s)
4.0 (s)
1
12 (s)
1
18 (s)
A/B SIM.
Default (unit)
2.5 (kHz)
75 (Hz)
20 (%)
ON
4 (s)
10 (min)
0.0 (s)
2.0 (s)
2.0 (s)
1
10 (s)
1
50 (s)
A/B SIM.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

100fx2c

Table of Contents