Accelerated Care Plus OMNISTIM FX2 PRO User Manual page 34

Electrotherapy system
Table of Contents

Advertisement

®
2
OMNISTIM
FX
PRO USER MANUAL
B. FAST CYCLE
Program simulates fast cycling, at 60 rpm, and can be used to retrain reciprocal movement of the L.E. with
or without resistance. Electrodes are applied to both quads while sitting with hips and knees flexed. Use a
lower extremity ergometer or pedal exerciser in combination with stimulation at mild to moderate
contraction intensity for closed chain functional training. Increase resistance as the patient progresses in the
rehab program.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
2
N
N
Y
1
C. SLOW WALK
Program simulates slow walking at a speed of 42 strides/minute. The program is used to re-educate gait
function. Electrodes are applied to both quads while sitting with hips and knees flexed. A light resistance
using weight cuffs or exercise band can be applied. The stimulation should cause a light to moderate
contraction. Transition to standing, with weight shifting activities then to walking in place in later rehab.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
2
N
N
Y
1
D. MEDIUM WALK
Program simulates medium-speed walking at a rate of 60 strides/minute. The program is used to re-educate
gait function. Electrodes are applied to both quads while sitting with hips and knees flexed. A light
resistance using weight cuffs or exercise band can be applied. The stimulation should cause a light to
moderate contraction. Transition to standing, with weight shifting activities then to walking in place in
later rehab.
User
Order in
Adjustable
Parameter
Parameter
Set Loop
(Y/N)
Y
2
N
N
Y
1
COPYRIGHT © 2004 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Parameter
Name
PHASE DURATION
CYCLE TIME
PULSE RATE
TREATMENT TIME
Parameter
Name
PHASE DURATION
CYCLE TIME
PULSE RATE
TREATMENT TIME
Parameter
Name
PHASE DURATION
CYCLE TIME
PULSE RATE
TREATMENT TIME
Adjustable
Range (unit)
40...100 (μs)
0...99 (min)
Adjustable
Range (unit)
40...100 (μs)
0...99 (min)
Adjustable
Range (unit)
40...100 (μs)
0...99 (min)
- 33 -
Default (unit)
70 (μs)
1.0 (s)
50 (Hz)
15 (min)
Default (unit)
70 (μs)
1.40 (s)
50 (Hz)
15 (min)
Default (unit)
70 (μs)
1.0 (s)
50 (Hz)
15 (min)