Agilent Technologies N6700 Series User Manual page 25

Modular power system
Hide thumbs Also See for N6700 Series:
Table of Contents

Advertisement

OUTPUT 2
+S
+
TWIST LEADS
Setting the Outputs
NOTE
Specifications for Series Operation
Load Transient
Recovery Time
Series N6700 User's Guide
OUTPUT 1
+S
+
-S
SENSE
JUMPERS
INSTALLED
+
LOAD
WITH LOCAL SENSING
First program the current limit of each output to the total desired
current limit point. Then program the voltage of each output so that
the sum of both voltages equals the total desired operating voltage.
The simplest way to accomplish this is to program each output to one
half of the total desired operating voltage.
The operating mode of each output channel is determined by the channel's
programmed settings, operating point, and load condition. Because these
conditions may change during parallel operation, the status annunciators on the
front panel will reflect these changes. This is normal. Momentary status
changes are also normal.
Specifications for outputs operating in series can be obtained from
the specifications for single outputs. Most specifications are
expressed as a constant or a percentage (or ppm) plus a constant.
For series operation, the percentage portion remains unchanged
while constant portions or any constants are changed as indicated.
Voltage
All series specifications referring to voltage are twice the single output
specification except for programming resolution, which is the same as for a
single output.
Current
All series specifications referring to current are the same as for a single
output except for CC load effect, CC load cross regulation, CC source effect,
and CC short term drift which are twice the current programming accuracy
(including the percentage portion).
Load transient specifications are typically twice the single output .
-S
Series Connections
OUTPUT 2
+S
+
-S
TWIST LEADS
+
LOAD
WITH REMOTE SENSING
OUTPUT 1
+S
+
-S
SENSE
JUMPER
INSTALLED
25