Agilent Technologies N6700 Series User Manual page 160

Modular power system
Hide thumbs Also See for N6700 Series:
Table of Contents

Advertisement

Error Messages
−400
−410
−420
−430
−440
202
202
202
202
202
160
Query Errors (these errors set Standard Event Status register bit #2)
Query Error
Generic error query
Query INTERRUPTED
A condition causing an interrupted query error occurred.
Query UNTERMINATED
A condition causing an unterminated query error occurred.
Query DEADLOCKED
A condition causing a deadlocked query error occurred.
Query UNTERMINATED after indefinite response
A query was received in the same program message after a query indicating an
indefinite response was executed.
Selftest Errors (these errors set Standard Event Status register bit #3)
Selftest Fail Aux Adc 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Auxiliary ADC failed. n1 and n2 are the expected limits. n3 is the measured value. n4 is the
channel location of the failed module.
Selftest Fail DACs 0 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
Selftest Fail DACs 1 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at zero; current DAC is at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
Selftest Fail DACs 2 expected <n1> to <n2>, measured <n3>, chan <n4>
Voltage DAC is at full scale; current DAC is at zero. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
Selftest Fail DACs 3 expected <n1> to <n2>, measured <n3>, chan <n4>
Both voltage and current DACs are at full scale. n1 and n2 are the expected limits. n3 is the
measured value. n4 is the channel location of the failed module.
Series N6700 User's Guide