STATus:PRESet
STATus:OPERation[:EVENt]? (@<chanlist>)
Bit Position
Bit Value
Bit Name
WTG-tran = The transient system is waiting for a trigger.
WTG-meas = The measurement system is waiting for
STATus:OPERation:CONDition? (@<chanlist>)
STATus:OPERation:ENABle <value>, (@<chanlist>)
STATus:OPERation:ENABle? (@<chanlist>)
Series N6700 User's Guide
This command sets all defined bits in the Status system's PTR
registers and clears all bits in the NTR and Enable registers.
Operation Register
STAT:OPER:ENAB
STAT:OPER:NTR
STAT:OPER:PTR
This query returns the value of the Operation Event register. The
Event register is a read-only register, which stores (latches) all
events that are passed by the Operation NTR and/or PTR filter.
Reading the Operation Event register clears it. The bit configuration
of the Operation status registers is as follows:
15-5
4
16
WTG-tran
a trigger
This query returns the value of the Operation Condition register.
That is a read-only register, which holds the live (unlatched)
operational status of the power system.
This command and its query set and read the value of the
Operational Enable register. This register is a mask for enabling
specific bits from the Operation Event register to set the operation
summary bit (OPER) of the Status Byte register. This bit (bit 7) is the
logical OR of all the Operatonal Event register bits that are enabled
by the Status Operation Enable register.
Questionable Register
STAT:QUES:ENAB
STAT:QUES:NTR
STAT:QUES:PTR
3
2
8
4
WTG-meas
OFF
OFF = The output is programmed off
CC = The output is in constant current
CV = The output is in constant voltage
Status Subsystem
Preset Settings
0 all bits disabled
0 all bits disabled
31 all defined bits enabled
3647 all defined bits enabled
1
0
2
1
CC
CV
89