Series N6700 User's Guide
Agilent Models N6751A/N6752A and N6761A/N6762A (continued)
Maximum Up-programming Time with full resistive load:
(time f
rom 10% to 90% of total voltage excursion)
Voltage setting from 0 V to 10 V
Voltage setting from 0 V to 50 V
Maximum Up-programming Settling Time with full resistive load:
(time from start of voltage change to within 50mV of final value)
Voltage setting from 0 V to 10 V
Voltage setting from 0 V to 50 V
Maximum Down-programming Time with no load:
(time f
rom start of voltage change to output voltage < 0.5 V)
Voltage setting from 10 V to 0 V
Voltage setting from 50 V to 0 V
Maximum Down-programming Settling Time with no load:
(time f
rom start of voltage change to output voltage within 50mV of final value)
Voltage setting from 10 V to 0 V
Voltage setting from 50 V to 0 V
Down-programming Time with 1000 µF load:
(time f
rom start of voltage change to output voltage < 0.5 V)
Voltage setting from 10 V to 0 V
Voltage setting from 50 V to 0 V
Down-programming Capability:
Continuous power
Peak current
Remote Sense Capability:
Series and Parallel Operation:
Load Cross Regulation:
Voltage, from no load to full load
Current, from no load to full load
2
Modules can discharge a 1000µF capacitor from 50V to 0V at a rate of 4 times/second.
Supplemental Characteristics
N6751A / N6752A
0.2 ms
1.5 ms
0.5 ms
4.0 ms
0.3 ms
1.3 ms
0.45 ms
1.4 ms
NOTE 2
2.1 ms
11 ms
7 W
7 A
Outputs can maintain specifications with up to a 1-volt
drop per load lead.
Similarly rated outputs can be operated directly in
parallel or can be connected for straight series
operation. Auto-series and auto-parallel operation is not
available.
1 mV
1 mA
N6761A / N6762A
0.6 ms
2.2 ms
0.9 ms
4.0 ms
0.3 ms
1.3 ms
0.45 ms
1.4 ms
4.5 ms
23 ms
7 W
3.8 A
0.5 mV
5 µA
111