3.1.3 Installation of ISC40G(S)-GS
with flange adapters
The ISC40G(S)-GS sensor is designed for
sanitary applications. For these applications
special process connections are necessary.

Mounting procedure:

• Screw the tube completely in the
stainless steel nut.
• Thread the sensor cable through
the flange adapter parts in the right
sequence.
• Screw the tube hand tight into the flange,
a mechanical stop will be felt.
• Tighten the plastic nut onto the sensor;
screw the plastic nut completely tight.
• Tighten the stainless steel nut for locking
the sensor.
Sensor is installed correctly if the flats on
the sensor are aligned perpendicular on the
process flow.
Tuchenhage
19 (0.75")
Ø71 (2.80")
/SFT
Fig 11: Option /SFT, /STC1, /STC2
Sanitary 2" Tri-clamp
O-ring
18.72 x 2.62
/STC1
plastic nut
tube
nut
ring
O-ring 18.7 x 2.6
ISC40G(S)-GS
Fig 10: Installation /SFT
Tri-clamp complete
Sanitary 2"
O-ring
18.72 x 2.62
17
Tuchenhagen
Flange
Check the O-ring
ø 60 x 3 for correct
mounting.
O-ring
18.72 x 2.62
2" clamp steal
/STC2