General Specifications; Measuring Elements; Materials; Functional Specifications (At 25°C) - YOKOGAWA ISC40G(S) User Manual

Table of Contents

Advertisement

2 GENERAL SPECIFICATIONS

The ISC40 inductive conductivity sensor is suitable for use with the Yokogawa inductive
conductivity analyzers.

2.1 Measuring elements

2.2 Materials

Wetted parts sensor
Body ISC40*-G*
Body ISC40*-T*
Non-wetted parts sensor
Sealing gasket
Thread part
Options for sensor
All options except /TFD : AISI 316 SS and O-ring material as wetted part
/TFD
2.3 Functional specifications (at 25°C)
Temperature element : Pt1000 to IEC 751
Installation factor
O n l y f o r P E E K
1.30
1.25
1.20
1.15
non conductive
piping
1.10
1,05
1.00
0.95
conductive
piping
0.90
0
10
20
Fig 1: Actual installation factor as function of spacing around the sensor
Note: The ISC40 temperature sensor is designed for cell compensation and for indication.
It is NOT designed for process temperature control.

2.4 Dynamic specifications

Response time
conductivity
: Toroïds with high permeability magnetic material
Pt1000 or 30k temperature element
: 30% glass filled PEEK, FDA approved
: PFA, FDA approved, PIM regulation 10/2011 approved
: Viton
: AISI 316 SS
: AISI 316 SS as non-wetted part
TFM and Kalrez as wetted part
30k NTC
: 1.88 cm-1 nominal for PEEK sensor
3.00 cm-1 nominal for PFA sensor
Actual installation can change this factor.
If there is less than 25mm spacing between sensor and holder,
in-situ calibration is necessary to meet the specified accuracies
(see fig. 1)
D
D i n m i l l i m e t e r s
30
40
50
: t
< 5 min. for PEEK sensor
90
:
t
< 10 min. for PFA sensor
90
9
IM 12D08J02-E-E