Control Drawing FM
The ISC40S sensor shall be installed with
one of the Yokogawa transmitters model:
• ISC202S • IC200S • FLXA21
with following parameters:
ISC202S
IC200S
Uo
14.4 V
Io
88 mA
162 mA
Po
317 mW
Lo
4.5 mH
Co
600 nF
or
To a FM approved intrinsically safe
apparatus meeting the entity parameters of
the ISC40S: Uo ≤ 19.1V
Io ≤ 170mA
Po ≤ 0.8W
Co ≥ Ci + C(cable)
Lo ≥ Li + L(cable)
The effective inductive capacitance Ci and the
effective induced inductance Li of the sensor
depends only upon the properties and the
length of the connected cable (max 50m).
When installing this equipment, follow the
manufacturer's control drawing. Installing
should be in accordance with ANSI/ISA
RP 12.06.01 "Installation of Intrisically
Safe Systems for Hazardous (Classified)
Locations" and the National Electrical Code
(ANSI/NFPA 70).
Warning: To prevent ignition of flammable or combustible atmospheres, disconnect power
before servicing or read, understand and adhere to the manufacturer's live maintenance
procedures.
ISC40-
-
-
Temperature Element
Plastic and adaption code
FLXA21
19.1 V
11.76 V
60.6 mA
178 mW
800µH
8 mH
254nF
100 nF
Cable length
in meters, any number between 1 and 50 meters
T1 PT1000
T3 30k thermistor
GG Glass filled PEEK, general model
GR Glass filled PEEK, retractable model
GS Glass filled PEEK, shaft model
TG PFA, general model
Fig 3: FF1-K1244QX
Control Drawing FM
13
Ta 85ºC/85ºC/40ºC
IM 12D08J02-E-E