3.4 Mounting FU20-FTS/MTS using
quick removal adapters
The FU20-FTS/MTS can also be mounted
using one of the optional quick-removal
adapters (/NSS, /NTI, /BSS, /BTI, see
Figure 5). For a detailed description of
these adapters see Sections 4 and 7 of
this Instruction Manual.
1 Apply Teflon tape to the threaded end of
this adapter;
2 Install the adapter in the process
connection. Tighten the adapter using a
wrench on the adapter flats.
1
3
5
Figure 5: Mounting of sensor with option /NSS, /NTI, /BSS or /BTI
IM 12B06J03-05EN-P
3 Apply Teflon tape to the appropriate
threaded end of the sensor;
4 Place the O-ring and screw the mating
part of the adapter on the sensor;
5 Mount the sensor in the adapter, making
sure that the O-ring seals properly;
6 Hand-tighten the adapter nut.
Other mounting examples of the FU20-
FTS/ MTS are given in Figure 6 and 7.
Note: DO NOT over tighten the adapter to
prevent damage.
The maximum Applicable torque is 8 Nm.