10
2.1.7. Shipping details
Package size (LxWxH)
Package weight
2.1.8. Environmental conditions
Storage temperature
2.1.9. Mechanical specifications
Surface roughness
2.2. Installation of SC4A sensors
For optimum measurement results, the SC4A sensor should be installed in a location that
offers an acceptable representation of the process composition and DOES NOT exceed the
specifications of the sensor.
2.2.1. Typical installation SC4A-*-AD with standard options
The SC4A-*-AD sensor can be installed in the process using optional ¾" NPT adapters. These
adapters are available in Stainless Steel (/PS) and in PVDF (/PF); see figure 3 for the mounting
sequence. The sensor can also be installed in an optional Stainless Steel flow fitting (/FF) using
option /PS, see figure 4 for details.
Fig. 3 Mounted sensor with the option /PS and /PF
IM 12D07J01-11E-E
SC4A-*-AD
SC4A-E-SA (SB, SC) : 0.5 µm
SC4A-*-PR
: 220 x 220 x 90 mm (8.7 x 8.7 x 3.6 inch)
: 0.5 to1.7 kg (1.1 to 3.8 lbs), depends on
sensor type and cable length
: -30°C to 50°C (-22°F to 122°F)
: 0.8 µm
: 0.8 µm
3/4" NPT
Ø30 (1.18")
1/4" NPT
Fig. 4 Dimensional drawing /FF
units in mm (inch)
1/4" NPT
30 (1.18")