Overview; Calibration (A And D Output) - Emerson Rosemount 1152 Alphaline Reference Manual

Nuclear pressure
Table of Contents

Advertisement

Reference Manual
00809-0100-4235, Rev BA
April 2007
Section 3

OVERVIEW

CALIBRATION (A AND D
OUTPUT)
www.emersonprocess.com/rosemount/nuclear
Calibration
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Calibration (A and D Output) . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Calibration (E, N, and L Output) . . . . . . . . . . . . . . . . . . . . page 3-4
Correction For High Line Pressure (All Output Codes) . page 3-11
Each transmitter is factory calibrated to the range specified by the customer.
This section contains the following transmitter calibration information:
Calibration—A and D Output
Span Adjustment (A and D Output)
Zero Adjustment (A and D Output)
Calibration Procedure—Zero and Span Adjustment (A and D Output)
Linearity Adjustment (A and D Output)
Damping Adjustment (D Output Only)
Calibration—E, N, and L Output
Span Adjustment (E, N, and L Output)
Zero Adjustment (E, N, and L Output)
Calibration Procedure—Zero and Span Adjustment (E, N, and L
Output)
Linearity Adjustment (E, N, and L Output)
Damping Adjustment (E, N, and L Output)
Correction for High Line Pressure—All Output Codes
Span
Zero
(See Nuclear Regulatory Commission IE Bulletin No. 80-16, June 27, 1980.)
The Rosemount 1152DP, HP, GP, and AP Transmitters are factory calibrated
to the range shown on the nameplate. This range may be changed within the
limits of the transmitter. Zero may also be adjusted to elevate (for all models
except Rosemount 1152AP) or suppress (for all models). The span and zero
adjustments are external and located under the nameplate.