Analog Receive (cont'd)
Name: Analog RX Test (Low Gain)
Description: Signal path test of the GRX boards with a TGC set to low gain.
Run Time: 00:02
If this test failed:
1.) Swap DRX boards to determine if the failure point changes as the boards are moved.
NOTE:
A block of 8 dead channels indicates a failed DRX board.
2.) Replace any failed boards.
3.) Replace GRX boards (these cannot be swapped as one is 64 channel and one is 128 channel).
7-5-12-9

Analog CW

NOTE:
Disconnect all probes before running diagnostics. Only connect a probe or test connector if
instructed for a specific test.
Figure 7-58 Analog CW and Patient I/O Service Diagnostics Options