Details Link On Diagnostic Window - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-5-7
Diagnostics Window Overview (cont'd)
11.)Select the Loop Order:
-
-
Some diagnostic windows that verify the tolerance of known values compile the test data and display it
in a table. You can access this data through the Details links.
12.)Click the Details icon to open the Details window.
13.)Click the See Details link to compile the Details data.
14.)The table displays the compiled detail data.
7 - 32
PROPRIETARY TO GE
. 7
EV
ABCABCABC = Runs the selected diagnostics as one group and runs each group for the
number of loops specified.
AAABBBCCC = Runs the selected diagnostics one at a time for the maxi m um loop value then
proceeds to the next diagnostic.
Figure 7-28 Advanced Features of Service Diagnostic Windows
Figure 7-29 Details Link on Diagnostic Window
Section 7-5 - Common Service Desktop
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL