AGMS – Solid State Multi Sensors
AGMS – M
For use with Accu-Gold, the AGMS-M is a Solid-State Dose multisensor for Mammographic Range
Measurements. It measures dose, dose-rate, time, kVp with FFT analysis, Flash HVL, and beam
filtration. It features Flash correction of dose for beam quality.
Type
Mammo
(AGMS-M)
Mammo (AGMS-M)
Reference Conditions:
2.2 mm polycarbonate compression paddle in beam for all measurements (kVp, dose, hvl).
Calibrated using large focal spot
Anode
/
Tube
Filter
Mo/Mo
General
Mo/Rh
General
Mo/Mo
GE
Mo/Rh
GE
Rh/Rh
GE
W/Ag
General
W/Rh
General
W/Al
General
Dose using
Dose Rate
AGDM
using
AGDM
150 nGy –
150 nGy/s –
>100
Gy,
350 mGy/s,
±5%
(17
±5%
(1.0
µR – >11
mR/min
kR, ±5%)
2.4 kR/min,
±5%)
USER GUIDE
kV ±2% or
Filt (um)
±0.7kV
whichever is
greater
21-49
16-44
21-49
6-42
22-48
16-44
22-48
6-42
25-48
12-80
20-38
12-80
20-38
12-78
20-50
200-990
Dose
using
Dose
Rate
AGDM+
using
AGDM+
80
nGy
-
80 nGy/s -
>100
Gy,
150 mGy/s
±5% (9.1 uR
±5%
(0.55
-
>11
kR,
mR/min - 1.0
±5%)
kR/min,
±5%)
hvl
(mm
AI)
±10% or ±0.05
mm
AI
whichever
is
greater
0.21-0.50
0.18-0.56
0.24-0.51
0.20-0.56
0.27-0.78
0.17-0.78
0.17-0.69
0.16-0.82
Time
1 ms – 300 s
± 0.1% +0.2
ms
71