Ma/Mas - Radcal ACCU-GOLD2 User Manual

Hide thumbs Also See for ACCU-GOLD2:
Table of Contents

Advertisement

Accu-Gold

mA/mAs

90M9-AG Invasive Sensor
The 90M9-AG is an invasive sensor. The mAs sensor connects to the 4-pin connector labeled mAs on
the digitizer using a 4-m cable and to the X-ray generator with safety shrouded banana jacks. It is
intended to be inserted in series with the X-ray generator return. The value of mAs is computed by
summing mA values during the part of the kV waveform (requires kV sensor in beam) defined by the
same thresholds that are applied to determine its width, generally 75% of kVp. Average mA is mAs
divided by this width.
Key features:
Automatic power control extends battery lifetime
Automatic zero
Measurement synchronized to kV waveform
.001 to 9999 mAs and 10 uA to 2A dynamic range with no range switching
Measures absolute value (no rectifier drop and optimum noise rejection)
mA-waveform
2.33-kHz -3-dB bandwidth
Specification / Technical Data
Full-scale
mA accuracy
mAs accuracy
Absolute value of bi-polar
Bandwidth
ON/OFF
Batteries
2000 mA or 9999 mA
0.2% of reading at dc plus resolution of +/-0.015 mA or 3-4
digits
(1-s pulse): 5 μAs or 0.2% of reading; resolution of 0.001 mAs
720 Hz waveform accurate to 2.5%
2.33 kHz, -3 dB
controlled from 9096 or 4086 control unit (ON only when a kV
function is active)
1 x 9.0 V Alkaline