Accu-Gold
13) Delete the active measurement.
14) Undo deleted measurement.
15) The Comments pane can be used to copy, cut and paste comments. Comments can be saved in
the
session along with any measurements and can optionally be carried over to the next
measurement. The comments pane can be accessed from all views.
16) Displays the calibration modality. Diagnostic or mammographic with anode/filter type.
17) Allows you to change settings. The available settings are:
Note: Settings available will depend on the sensor attached and therefore will not show up until a sensor
is attached and enabled by pressing Connect.
Device Settings Menu
Trigger Source
The upper row of the Settings area is populated by the sensors that are connected (AGMS, DD and/or
ICH). When multiple sensors are connected, the user can choose the trigger source.
Trigger Level
Std - The default setting. Std" trigger sensitivity is recommended.
Low - Select if Std is not low enough. "Low" may allow smaller signals to be captured, but may also
result in false triggers
5
•
NOTE: Do not make sensitive measurements when the computer is hooked up to a charger
unless the charger has a 3-wire AC cord (and the AC outlet is properly grounded). See AG-GND
in appendix.
5
. If noise or false triggers prevent reliable ion chamber measurements and
USER GUIDE
13