Measurements; License Files; Using The Agms-Dm+ Sensor; Taking A Measurement For Mammographic Sensors - Radcal ACCU-GOLD2 User Manual

Hide thumbs Also See for ACCU-GOLD2:
Table of Contents

Advertisement

Accu-Gold
Filt:
Mode:
Notes:
Copy – Highlight the text to be copied and click on the copy icon. This will copy the Comment text
to the clipboard for use in other applications.
Paste – The contents of the clipboard will be pasted in the comments pane starting at the location
of the cursor.
Carry-Over – When the push-pin is black, the comments pane information will be copied to the
comments pane of the subsequent measurement. This can be very useful when conditions are not being
changed between measurements.

Measurements

License Files

When the USB flash drive is used to install the software that was supplied with your system, any license
file that is required will be automatically installed. For Legacy AGMS sensors, license files contain
calibration tables that match the sensor's serial number. If you are using a Legacy multi-sensor (AGMS-
D or AGMS-M) and have not installed the license file or if they have been recalibration you can use the
Main Menu dropdown and select Licenses/Import license. If the AG software cannot locate a matching
license file, it will provide a dialog box asking you to locate it. Plus, sensors have the calibration files built
into the sensors.
Custom calibrations, if not installed as part of software setup, would also be added this way.
NOTE: The CPC date (Calibration Parameters Change date) is saved with the sensor information for
every measurement. When a sensor is sent in for annual calibration, the sensor's calibration is verified
and the date is changed only if the calibration parameters need to be changed. If the sensor's
parameters are within limits, the original CPC date is not changed, only the certification label is updated.

Using the AGMS-DM+ sensor

Once the sensor is connected, select the appropriate modality from the pop-up list in the bottom right
corner. This pop-up menu appears when multi-sensor is connected. If it is a DM+ sensor, the default will
be Diagnostic.

Taking a Measurement for Mammographic Sensors

Before making an exposure, make sure to choose the Anode/Filter combination that best represents the
machine being measured.
Radcal sensor calibrations used the following mammographic tubes and assume a 2.2 mm
polycarbonate paddle, or simulated paddle (Appendix G), are placed on top of the sensor:
Anode/Filter 
Mo‐Mo 
Mo‐Mo GE 
Mo Rh 
Mo Rh GE 
X‐Ray tube used 
Lorad M2 
GE DMR 
Lorad M2 
GE DMR 
USER GUIDE
21