AGMS – M (DISCONTINUED)
For use with Accu-Gold, the AGMS-M is a Solid-State Dose multisensor for Mammographic Range
Measurements. It measures dose, dose-rate, time, kVp with FFT analysis, Flash HVL, and beam filtration.
It features Flash correction of dose for beam quality.
Type
Anode / Filter
Mammo
Mo/Mo
(AGMS-M)
Mo/Rh
Mo/Mo
Mo/Rh
Rh/Rh
W/Ag
W/Rh
W/Al
Mammo
Dose
(AGMS-M)
using
AGDM
150nGy–
>100Gy,±5%
(17 R–
>11kR,±5%)
Reference Conditions:
2.2 mm polycarbonate compression paddle in beam for all measurements (kVp, dose, hvl).
Calibrated using large focal spot
Tube
kV
±2% or ±0.7kV
whichever
is greater
General
21-49
General
21-49
GE
22-48
GE
22-48
GE
25-48
General
20-38
General
20-38
General
20-50
Dose Rate
Dose
using
using
AGDM
AGDM+
150nGy/s–
80nGy-
350mGy/s,±5%
>100Gy,±5%
(1.0mR/min–
(9.1uR-
2.4kR/min,±5%)
>11kR,±5%)
Filt (um)
hvl (mm AI) ±10% or
±0.05 mm AI
whichever is greater
16-44
0.21-0.50
6-42
0.18-0.56
16-44
0.24-0.51
6-42
0.20-0.56
12-80
0.27-0.78
12-80
0.17-0.78
12-78
0.17-0.69
200-990
0.16-0.82
Dose Rate using
Time
AGDM+
80nGy/s-
150mGy/s,±5%
1 ms – 300s
(0.55mR/min-
±0.1%+0.2 ms
1.0kR/min,±5%)
80

Advertisement