Page 1
Model DAPCKP DAPcheck Plus DAP Analyzer Instruction Manual...
Page 3
Made in Germany for Radcal Corporation under the Following Patent: Pat No. 7,413,345 Radcal Corporation 426 West Duarte Road Monrovia, CA 91016-4591 USA USA (626) 357-7921 Fax USA (626) 357-8863 email Service@radcal.com www.radcal.com Radcal Part # MNL/DAPCKP 4077023 Rev:- Printed: Nov, 2012...
Page 4
Important. Throughout this manual the following words are given a specific meaning and the reader should be aware of their significance. Attention; Used where there is a risk of damaging the DAPcheck Plus incorrect handling or operation error. Note; Used where information is provided to aid better understanding of or its use.
General user information Attention. This section contains essential information that is required for the safe use of the DAPcheck Plus. Application The DAPcheck Plus has been designed for checking the calibration of dose area product meters and air kerma meters used in patient dosimetry. It can also be used to establish reference dose area product levels for specific examinations.
Disposal The DAPcheck Plus must be disposed of in accordance with the applicable national regulations or returned to the manufacturer. A WEEE manual is available from Radcal.
System description Introduction The DAPcheck Plus is designed to function as an external sensor for use with a Accu-Gold Digitizer. The Accu-Gold Excel software operates the DAPcheck Plus though a computer interface. (See separate software user manual). A support stand is provided to elevate the DAPcheck Plus above structures that might add scattered radiation to a primary beam measurement.
Preparations for operation Preparation The DAPcheck Plus should be carefully removed form its packing case and inspected. If any damage to the components is found, the supplier or manufacturer should be contacted. Damaged components must not be used. Power-on test Note.
Excel add-on. A quick start manual can be found under Manuals in the same folder. (Use of AccuGoldExcel requires that Microsoft Excel is already installed on your computer).To run the software, open the Accu-Gold Excel folder using the shortcut or navigate to ‘Program Files\Radcal\AccuGoldExcel’. A simple template for making single measurements is called: AccuGold-DAPCheckPlus...
Technical specifications Notes: All technical data is valid for the ambient conditions as defined in IEC 60580. Only values given with tolerance ranges or limits are guaranteed. All other values are for information only. Subject to change without prior notice. Accuracy ±...
Page 12
One meter disconnect cable with 2 Connection to Accu-Gold Digitizer connectors IP 41 Protection Class (acc. IEC 60529) 2.3 kg; 5.1 lbs. Weight 350 mm x 410 mm x 35 mm; Dimension (L x W x H)
Page 13
Support Elevation Stand Foam Elevation Support Stand fits inside the interior of the Carry Case lid...
Radcal if the repairs are warranty- applicable. This warranty excludes batteries. Radcal shall not be held liable for damages or delays caused by defects beyond making repairs or furnishing replacement parts, nor shall Radcal be liable for any defective material replaced without Radcal’s consent during the period of this warranty.
Page 16
Declaration of Conformity According to ISO/IEC 17050 and EN 45014 The Radcal Corporation declares, under our sole responsibility, that the model DAPcheck Plus Measurement System conforms to the following product specifications. EMC: EN50081-2/EN55011, Class A EN61000-4-2, EN61000-4-3, EN61000-4-8 IEC60601-1 IEC60601-1-2...
Need help?
Do you have a question about the DAPCKP and is the answer not in the manual?
Questions and answers