DC Voltage Measurement

Range
110 mV
1.1V
11V
110V
300V
Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/ C in the ranges -10 to 18 C and 28 to 50 C
Input Impedance: 5 M
Common Mode Error: 0.008% f.s./(Common Mode Volt)
Maximum Input Voltage: 300V rms
Resolution
1 V
10 V
100 V
1 mV
10 mV
Models 741B and 743B
% of Reading +% of Full Scale
1 Year
0.025% + 0.015%
0.025% + 0.005%
0.025% + 0.005%
0.05% + 0.005%
0.05% + 0.005%
Documenting Process Calibrator
Specifications
2 Year
0.05% + 0.015%
0.05% + 0.005%
0.05% + 0.005%
0.1% + 0.005%
0.1% + 0.005%