Quick Installation Guide
00825-0100-4648, Rev CC
February 2012
S
1
TEP
CONTINUED
5. Close the housing cover and tighten to safety specification. Always ensure a proper seal
by installing the electronics housing covers so that metal touches metal, but do not over
tighten.
6. Position the antenna such that it is vertical, either straight up or straight down. The
antenna should be approximately three ft. (0,91 m) from any large structures or
buildings, to allow clear communication to other devices.

Remote Mount

1. Install the sensor according to standard installation practices. Be sure to use an
approved thread sealant on all connections.
2. Run wiring (and conduit, if necessary) from the sensor to the 648.
3. Pull the wiring through the threaded conduit entry of the 648.
4. Attach the sensor wiring to the terminals as indicated on the wiring diagram.
5. Connect the Power Module.
NOTE:
Wireless devices should be powered up in order of proximity from the Wireless
Gateway, beginning with the closest device to the Gateway. This will result in a
simpler and faster network installation.
...
Possible antenna rotation shown.
Antenna rotation allows for best
installation practices in any
configuration.
Possible antenna rotation shown.
Antenna rotation allows best
practices for any configuration.