Braun Infusomat Space Instructions For Use Manual page 68

Automatic infusion system
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

Chapter 8
68
Volume increments
Time selection
Accuracy of set delivery rate
Administration Set Change Interval
Max. Volume in case of single fault
condition
Technical inspection (safety check)
Rate increments
Multiple lines connected to one
patient port
Accuracy of bolus infusion
KVO rate
Computer connection
for Service Program
Air detector
TECHNICAL DATA
0.1 - 99.99 ml in increments of 0.01 ml
100.0 – 999.0 ml in increments 0.1 ml
1000 – 9999 ml in increments 1 ml
00:01 – 99:59 h
± 5 % according to IEC/EN 60601-2-24
Pumping accuracy is maintained for a
minimum of 96 hours.
For incorrect dosages of 1.4 ml due to
malfunctions of the device the pump will
automatically shut off
Every 2 years
0.1 - 99.99 ml/h in increments of
0.01 ml/h
100.0 - 999.9 ml/h in increments of
0.1 ml/h
1000.0 - 1200 ml/h in increments of
1 ml/h
Connecting multiple infusion lines with
different flow rates may affect the rate
for all infusions past the point of
connection.
typically ± 5 % for a bolus volume > 1 ml
Delivery rate
10 ml/h: KVO rate 3 ml/h
Delivery rate < 10 ml/h: KVO rate 1 ml/h
Delivery rate < 1 ml/h: KVO rate =
programmed rate (default setting 0.1 ml/h)
USB connection in combination with
B. Braun interface lead CAN SP
(8713230) including electrical
insulation. Please pay attention to
safety notices.
The Infusomat® Space Volumetric
Infusion pump includes an ultrasonic air
detection sensor that is used to detect
air in the disposable set. Per the
requirements of the IEC 60601-2-24
(2nd Ed. 51.104 and 3rd Ed.
201.12.4.4.107) the software uses this
sensor to detect the accumulation of air
not to exceed 1 ml/15 minutes of any
air bubbles greater than or equal to
0.01 ml (10 uL). The air accumulating