Braun Infusomat Space Instructions For Use Manual page 39

Automatic infusion system
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

Chapter 1
• A second prompt requires responding yes with
• Respond "yes" or "no" to "Use dose rate calc?" (if configured). See Section 2.2.4.
Note:
• The pump requires user to enter 2 out of 3 of the following parameters: rate,
VTBI or time. When 2 are entered, the pump calculates the 3rd parameter.
SECondary, programmed bolus and manual bolus are also available in Basic mode, See
Section 1.4, 1.7, 1.8. The target symbol
other than rate, that was first set by the user. When a rate titration is made the value
with the
rate. As an example if a rate of 10ml/hr and time of 20hrs is programmed the VTBI is
calculated, when rate is titrated the VTBI is changed not the time. If the rate and VTBI
are initally programmed the time would change with a change in rate. The parameter
with the target symbol does not change during titration of either of the other 2 para-
meters.
1.) Enter VTBI and time: The infusion rate will be calculated and displayed on the
bottom of the display.
• Select VTBI with
• Enter VTBI with
• Select time with
• Enter time with
Rate titration will result in adjustment of time. Alternatively:
2.) Enter rate and VTBI: The infusion time will be calculated and displayed on the
bottom of the display.
Rate titration will result in re-calculation of time.
While under Basic Infusion, the pump is not using any drug library safety limits.
is not changed, rather the 3rd calculated value is adjusted for the new
t
and open with l.
q
and confirm with
t
and open with l.
q
and confirm with
u
to "Continue without limits".
appears next to the programmed value,
k
.
k
.
OPERATION
39