Braun Infusomat Space Instructions For Use Manual page 29

Automatic infusion system
Hide thumbs Also See for Infusomat Space:
Table of Contents

Advertisement

Chapter 1
"Back to PRIM" allows the pump to be set to require a manual switch back to PRIMary
in which case the pump stops and alarms when the SECondary is complete, or auto-
change whereby the pump converts to the PRIMary setting when the SECondary
dose/VTBI has been infused.
Note:
until changed.
• Select "New SECondary"
• Select the drug category and scroll to the desired drug name.
The Care Area and Patient Profile are retained from the PRIMary infusion. Select
infusion mode and concentration as done above whenprogramming a PRIMary
infusion.
Note:
appear in the drug list when in the SECondary menu.
Note:
A message will appear indicating SECondary has been disabled for the drug.
SECondary meds, which are administered intermittently, will most frequently be set up
for Dose over Time. SECondary meds set up as continuous are programmed as above
for PRIMary programming.
• Enter total dose and confirm with k.
• Enter total time and confirm with k.
Note:
• Press
secondary.
• Press
The method of switching back to PRIMary remains for all SECondary infusions
Medications may be set up in the drug library for SECondary only and will only
PRIMary medications may be set in drug library to prohibit SECondary infusions.
The pump calculates the rate and VTBI.
f
s
key, note message reminder to check bag height and unclamp
f
s
again to confirm message and start SECondary.
OPERATION
29