PUMP MENUS
Chapter 2
• Select patient parameter, if any, for dosing calculation. Choices are weight,
BSA, or none.
• Select the doserate units.
• Enter doserate using
q
arrow keys and confirm with k.
Confirm and start infusion.
• *Enter VTBI or Time.
• To exit Dose Rate Calculation: the pump must be stopped. Press the
c
Key
from Home screen and answer "yes" to "Exit and clear infusion?"
Assigning a basic or doserate calculation infusion to Drug Library:
An infusion started without using the drug library, either a basic or doserate
calculation, may be assigned to the drug library without stopping the infusion.
• Access the Home screen by pressing the
c
key.
• Scroll to and select Options.
• Scroll to and select "Assign to Drug Library".
• Program for the Drug Library following the same steps covered in Chapter
1.3 to program within the drug library, beginning with selecting the Care Unit.
46