Quick Start Guide
8.6
China
I3 China Intrinsic Safety
Certificate: GB3836.1-2010, GB3836.4-2010, IEC61241-0 - 2004,
Markings: Ex ib/ia IIB Gb T4
Special Condition for Safe Use (X):
1. The installation of the product shall provide a degree of protection IP6X according to
GB4208-2008, and in such a way that all metallic parts are integrated in the local
equipotential bonding.
8.7
Brazil
I2 INMETRO Intrinsic Safety
Certificate: UL-BR 15.0072X
Standards: ABNT NBR IEC 60079-0:2008 + Errata 1:2011,
Markings: Ex ib [ia] IIB T4 Gb (- 40 °C < T
Special Condition for Safe Use (X):
1. Dust application:
The installation of the 3D Solids Scanner or of the Antenna Unit of models providing
head separation in the wall to areas requiring EPL Da (Zone 20) equipment shall
provide a degree of protection IP6X according to ABNT NBR IEC 60529 and shall be
carried out in such a way, that all metallic parts are integrated in the local equipotential
bonding.
Manufacturer's technical information related to use of the 3D Solids Scanner in contact
with aggressive/corrosive media and to avoid any risk of mechanical impact shall be
observed.
32
GB12476.4-2010
Ex ibD/iaD 21/20 T110 °C
ABNT NBR IEC 60079-11:2009
Ex ib [ia] IIIC T110 °C Da/Db (- 40 °C < T
< + 85 °C)
a
< + 85 °C)
a
July 2017