On the Device menu, select Echo Curve Analysis. Then make sure the
check-box is checked, then select the Start button.
Once the Echo Curve Analysis is complete, the echo curve window will pop up.
This function is also available via Device > Echo Curve Analyze Viewer.
5.4
Performing False Echo Mapping
With this option it is possible to perform a false echo mapping on any of the
beams to ignore false echoes inside the vessel caused by internal objects or
other interferences.
1. On the Device menu, select Device False Echo Mapping.
2. Set the From and To distances to perform false echo mapping.
3. Select the Start Scanning button.