Manual Mode; Starting A Centrifugation Run; Interrupting A Centrifugation Run; Interrupting A Deceleration Process - Sigma 2-16P Operating Manual

Laboratory centrifuge
Hide thumbs Also See for 2-16P:
Table of Contents

Advertisement

Laboratory Centrifuge Sigma 2-16P
6 Using the centrifuge
Pos : 120 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontrol Uni versal /062- 0030-0020 M anueller Betri eb @ 28\mod_1405320526938_68.doc x @ 194439 @ 3 @ 1
6.3.2

Manual mode

Pos : 121 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontrol Uni versal /062- 0030-0020-0010 Starten ei ner Z entrifugati on @ 28\mod_1405320527735_68.doc x @ 194453 @ 4 @ 1
6.3.2.1

Starting a centrifugation run

Pos : 122 /010 Uni vers almodul e/
Leerzeil e 8pt @ 12\mod_1337063062114_0.doc x @ 65687 @ @ 1
Pos : 123 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontr ol Uni versal /062- 0030-0020-0020 U nterbr ec hen einer Z entrifug ation @ 28\mod_1405320528561_68.doc x @ 194467 @ 4 @ 1
6.3.2.2

Interrupting a centrifugation run

Pos : 124 /010 Uni vers almodul e/
Leerzeil e 8pt @ 12\mod_1337063062114_0.doc x @ 65687 @ @ 1
Pos : 125 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontrol Uni versal /062- 0030-0020-0030 U nterbr ec hen eines Brems vorgangs @ 28\mod_1405320529412_68.doc x @ 194481 @ 4 @ 1
6.3.2.3

Interrupting a deceleration process

Pos : 126 /010 Uni vers almodul e/
Leerzeil e 8pt @ 12\mod_1337063062114_0.doc x @ 65687 @ @ 1
Pos : 127 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontrol Uni versal /062- 0030-0020-0040 Aus wahl, Anzeig e und Änderung von D aten @ 28\mod_1405320530250_68.doc x @ 194495 @ 4 @ 1
6.3.2.4

Selection, display, and modification of data

Pos : 128 /010 Uni vers almodul e/
Leerzeil e 8pt @ 12\mod_1337063062114_0.doc x @ 65687 @ @ 1
Pos : 129 /100 Sig ma/100 BA Z entrifugen Sig ma (Standardmodul e)/062 Betrieb_Spi nc ontr ol Eas y+U ni versal/Spinc ontrol Uni versal /062- 0030-0020-0050 Laufzei t - ohne T emp @ 28\mod_1405320531525_68.doc x @ 194509 @ 4 @ 1
6.3.2.5

Runtime

NOTE
34 / 72
The centrifuge is ready for operation when the start key is illuminated.
Press the start key in order to start a centrifugation run.
Press the stop key in order to interrupt a centrifugation run.
The centrifugation run will be terminated prematurely.
Press the start key during a deceleration process in order to interrupt it
and to restart the centrifuge.
Turn the left function knob in order to select a field. The selected field
will be preceded by the indication "set", "run", or "save".
Press the left function knob. The indication ("set", "run", or "save")
flashes and the modification mode is active.
Turn the right function knob in order to modify the set value of the
selected field.
Press the fright unction knob again to confirm the input and to quit the
modification mode
If the input is not confirmed, the value will be automatically reset to the
last setting.
The runtime is displayed in the left section of the display and can be set at
different intervals in a range from 10 seconds to 11 hours and 59 minutes.
The parameter runtime can be changed during the centrifugation.
If the runtime is changed during an active centrifugation run, the time that
has already elapsed will not be taken into consideration. The centrifuge will
perform a complete run with the new runtime.
Fig. 10: Setting the runtime
Translation from the original operating manual, part no. 0700602
Version 04/2007, Rev. 2.15 of 15/09/2017 • sb