Series 2600B System SourceMeter® Instrument Reference Manual
Example
smua.measure.rangev = 0.5
Also see
Range
(on page 2-81)
reset()
(on page 7-173)
setup.recall()
smuX.measure.autorangeY
smuX.reset()
smuX.source.rangeY
smuX.measure.rel.enableY
This attribute turns relative measurements on or off.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
relEnable = smuX.measure.rel.enableY
smuX.measure.rel.enableY = relEnable
relEnable
X
Y
Details
This attribute enables or disables relative measurements. When relative measurements are enabled,
all subsequent measured readings are offset by the relative offset value specified by
smuX.measure.rel.levelY. Each returned measured relative reading is the result of the following
calculation:
Relative reading = Actual measured reading − Relative offset value
Example
smua.measure.rel.enablev = smua.REL_ON
Also see
Relative offset
reset()
(on page 7-173)
setup.recall()
smuX.measure.rel.levelY
smuX.reset()
2600BS-901-01 Rev. C / August 2016
(on page 7-194)
(on page 7-214)
(on page 7-232)
(on page 7-247)
Affected by
Instrument reset
SMU reset
Recall setup
Relative measurement control. Set relEnable to one of the following values:
0 or smuX.REL_OFF: Disables relative measurements
1 or smuX.REL_ON: Enables relative measurements
Source-measure unit (SMU) channel (for example, smua.measure.rel.enablev
applies to SMU channel A)
SMU measurement function (v = voltage, i = current, r = resistance, p = power)
(on page 3-1)
(on page 7-194)
(on page 7-228)
(on page 7-232)
Section 7: TSP command reference
Selects the 1 V measurement range for SMU channel A.
Where saved
Default value
Not saved
0 (smuX.REL_OFF)
Enables relative voltage measurements for SMU
channel A.
7-227