Smux.source.compliance - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Section 7: TSP command reference
Also see
Calibration
(on page B-1)
smuX.cal.restore()
smuX.cal.save()
smuX.cal.unlock()
smuX.measure.calibrateY()

smuX.source.compliance

This attribute contains the state of source compliance.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
compliance = smuX.source.compliance
compliance
X
Details
This read-only attribute contains the source compliance state (true or false).
true indicates that the limit function is in control of the source (source in compliance).
false indicates that the source function is in control of the output (source not in compliance).
Writing to this attribute generates an error.
Reading this attribute also updates the status model and the front panel with generated compliance
information. See Current Limit (ILMT) shown in the status model diagram for the
registers
(on page E-8). The Voltage Limit (VLMT) is not shown in the status model diagram for the
Measurement event registers
Example
compliance = smua.source.compliance
print(compliance)
Also see
smuX.source.limitY
7-236
(on page 7-203)
(on page 7-204)
(on page 7-205)
(on page 7-216)
Affected by
Not applicable
The state of source compliance
Source-measure unit (SMU) channel (for example, smua.source.compliance
applies to SMU channel A)
(on page E-8), but is similar to the Current Limit (ILMT).
(on page 7-240)
Series 2600B System SourceMeter® Instrument Reference Manual
Where saved
Not saved
Default value
Not applicable
Measurement event
Reads the source compliance state
for SMU channel A.
Output:
true
This output indicates that a
configured limit has been reached
(voltage, current, or power limit).
2600BS-901-01 Rev. C / August 2016