Tspnet.disconnect() - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual
Example 2
instrumentID = tspnet.connect("192.0.2.1", 1394, "*rst\r\n")
if instrumentID then
-- Use instrumentID as needed here
tspnet.disconnect(instrumentID)
end
Also see
localnode.prompts
localnode.showerrors
tspnet.tsp.abortonconnect

tspnet.disconnect()

tspnet.disconnect()
This function disconnects a specified TSP-Net session.
Type
TSP-Link accessible
Function
No
Usage
tspnet.disconnect(connectionID)
connectionID
Details
This function disconnects the two devices by closing the connection. The connectionID is the
session handle returned by tspnet.connect().
For TSP-enabled devices, this aborts any remotely running commands or scripts.
Example
testID = tspnet.connect("192.0.2.0")
-- Use the connection
tspnet.disconnect(testID)
Also see
tspnet.connect()
2600BS-901-01 Rev. C / August 2016
(on page 7-154)
(on page 7-158)
(on page 7-407)
(on page 7-401)
Affected by
The connection ID returned from tspnet.connect()
(on page 7-400)
Section 7: TSP command reference
Where saved
Default value
Create a TSP-Net session.
Close the session.
Connect to a device that is
not TSP-enabled.
7-401

Advertisement