Status.request_Enable - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Section 7: TSP command reference
As an example, to set bit B1 of the questionable status unstable output summary enable register, set
status.questionable.instrument.enable = status.questionable.instrument.SMUA
.
In addition to the above constants, questionableRegister can be set to the numeric equivalent of
the bit to set. To set more than one bit of the register, set operationRegister to the sum of their
decimal weights. For example, to set bits B1 and B2, set questionableRegister to 6 (which is
the sum of 2 + 4).
Example
status.questionable.unstable_output.enable =
status.questionable.unstable_output.SMUA
Also see
Questionable Status Registers
status.questionable.*

status.request_enable

This attribute stores the service request (SRQ) enable register.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
requestSRQEnableRegister = status.request_enable
status.request_enable = requestSRQEnableRegister
requestSRQEnableRegister
Details
This attribute is used to read or write to the service request enable register. Reading the service
request enable register returns a value. The binary equivalent of the value of this attribute indicates
which register bits are set. In the binary equivalent, the least significant bit is bit B0, and the most
significant bit is bit B7. For example, if a value of 1.29000e+02 (which is 129) is read as the value of
this register, the binary equivalent is 1000 0001. This value indicates that bit B0 and bit B7 are set.
B7
B6
**
>
1
0
* Least significant bit
** Most significant bit
7-340
Series 2600B System SourceMeter® Instrument Reference Manual
(on page E-13)
(on page 7-330)
Affected by
Status reset
The status of the service request (SRQ) enable register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate
various bit settings
B5
B4
B3
>
>
>
0
0
0
Uses a constant to set the SMU A
bit in the questionable status
unstable output summary enable
register bit.
Where saved
Default value
Not saved
0
B2
B1
B0
>
>
*
0
0
1
2600BS-901-01 Rev. C / August 2016