Return Range
Return Range
Keysight M8070A Programming Guide
:PLUGin:ERATio:ACQuisition:TCLevel[?]
Syntax
:PLUGin:ERATio:ACQuisition:TCLevel 'identifier', <NRf>
:PLUGin:ERATio:ACQuisition:TCLevel? 'identifier'
Input
'identifier': Specify the measurement name.
Parameters
<NRf>: Set target confidence level.
0.1% to 99.9%
Description
This command sets the target confidence level when accumulation end is
set to pass/fail (in percent) for the addressed measurement name
identifier. The query returns the current target confidence level.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:ERAT:ACQ:TCL 'MyMeasurement', 95
:PLUGin:ERATio:ACQuisition:TERatio[?]
Syntax
:PLUGin:ERATio:ACQuisition:TERatio 'identifier', <NRf>
:PLUGin:ERATio:ACQuisition:TERatio? 'identifier'
Input
'identifier': Specify the measurement name.
Parameters
<NRf>: Set target error ratio.
1E-18 to 1E-3
Description
This command sets the target error ratio for the addressed measurement
name identifier. The query returns the current target error ratio value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:ERAT:ACQ:TER 'MyMeasurement', 1e-09