5
SCPI Command Reference
Return Range
154
[:SOURce]:JITTer:HFRequency:PMMode[?]
Syntax
[:SOURce]:JITTer:HFRequency:PMMode 'identifier', <UI|TIME>
[:SOURce]:JITTer:HFRequency:PMMode? 'identifier'
Input
'identifier': 'M*.DataOut1' or 'M*.DataOut2'
Parameters
<UI|TIME>: Specify unit interval or time.
UI|TIME
Description
The internal delay line is optimized for either 'UI' (1UI; the maximum of the
delay line is not used) or for 'TIME' rather than the whole range of the
delay line, which can be used independent of the data rate. PMMode
stands for Phase Modulation Mode.
This SCPI is applicable for M8041A, M8051A and M8045A.
Example
:JITT:HFR:PMM 'M1.DataOut2', UI