Summary of Contents for Keysight Technologies M9405A
Page 1
Keysight Technologies M9405A NOTICE: This document contains references to Agilent Technologies. Agilent’s former Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. User's and Service Guide...
Description Congratulations on your purchase from Keysight Technologies, Inc. The M9405A is a 1-slot, 1U PXIe Amplifier. The M9405A operates over a frequency range of 300 kHz to 26.5 or 50 GHz. To get the most out of your purchase, it is recommended that this User’s and Service Guide be read carefully and completely.
M9405A Technical Specification and Characteristics Technical Specification and Characteristics Definitions Specifications describe the warranted performance of calibrated instruments. Data represented in this document are specifications unless otherwise noted under the following conditions. • Full temperature range = Individual module temperature of as reported by the ...
M9405A Electrostatic Discharge Protection Electrostatic Discharge Protection Electrostatic discharge (ESD) can damage or destroy electronic components. The module is shipped in materials which prevent damage from static, and should only be removed from the packaging in an anti-static area ensuring that the correct anti-static precautions are taken.
M9405A General Characteristics General Characteristics Table 2 Nominal Performance Frequency Range 300 kHz to 50 GHz 0.1 dB 1 dB 1 GHz -22 -18 Input Power at Compression (dBm) 10 GHz -22 -13 (0.1 dB and 1 dB compression) 26.5 GHz...
M9405A General Characteristics Front Panel Features Figure 2 M9405A Front Panel Status LED Access Ports Latch Status LED • Green, the status LED indicates the chassis is powered on and communicating with the module when the soft front panel is opened (SFP).
M9405A Install the Software Install the Software Table 3 System and Hardware Requirements Topic Requirements System Requirements Window XP, Windows Vista®, SP1 and SP2 (32-bit and 64-bit), Operating System Service Pack 3 Windows 7 (32 bit and 64 bit) 600 MHz or higher required ...
It is also available on the Internet at: http://www.keysight.com/find/iosuite • Instrument software, which includes the soft front panel (SFP), device driver (IVI-COM) and documentation for the M9405A. This software CD is included with your shipment (M9499-10002). Use the following procedure to install the software: 1.
Install the Software Install the M9405A The M9405A can be used in a chassis with a PXIe or PXI-H chassis peripheral slot. The M9405A can also be installed in any hybrid PXI slot marked with a peripheral slot compatibility image (the letter "H" and a solid circle containing the slot number).
Page 14
M9405A Install the Software 5. Secure the front panel to the chassis using the front panel mounting screws. Performance may suffer if the screws are not tightened properly. 6. Verify that the PXI chassis fans are operable and free of dust and other contaminants that may restrict air flow.
M9405A Operational Check Procedure M9405A Operational Check Proced ure The operational check measures the signal paths from the RF Input of the M9405A module. The following procedure verifies that the connectors and circuitry are operating correctly. Do not exceed the maximum input power level of +20 dBm.
M9405A Operational Check Procedure Equipment Test Configuration Option II (Signal Source & Power Meter) 1. Adjust the Signal Source Stimulus Power for the M9405A to -30 dBm. 2. Set the Signal Source to a CW Frequency of 50 MHz. 3. Perform a Power Meter calibration.
M9405A Service Service If a Problem is Found If a problem is found, complete the following procedures: 1. Verify that the relevant hardware is turned on. 2. Verify that the stimulus signal is set to the proper power/frequency and that all cables are properly connected.
11.) Program the replacement module with the serial number from the defective module. a. If you don’t already have the Keysight M9405A Simple Module SFP, download it from http://www.keysight.com/find/M9405A (from this site, select Technical Support > Drivers, Firmware & Software), and install it on your computer or embedded controller.
Page 19
Write the following information on a tag and attach it to the malfunctioning equipment: • Name and address of owner. A P.O. box is not acceptable as a return address. • Product model number (for example, M9405A). • Product serial number. The serial number label is located on the side panel of the module.
M9405A General Safety and Information General Safety and Information Introd uction Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with accepted industry standards, and has been supplied in a safe condition.
M9405A Regulatory Information Regulatory Information Listed below are definitions for the markings that may be found on the product. Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
M9405A Regulatory Information EMC Compliance Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates editions are cited in the Declaration of Conformity): • IEC/EN 61326-1 • CISPR Pub 11 Group 1, Class A •...
Keysight Support, Services, and Assistance Service and Support Options There are many other repair and calibration options available from the Keysight Technologies support organization. These options cover a range of service agreements with varying response times. Contact Keysight for additional information on available service agreements for this product.