Return Range
Return Range
Keysight M8070A Programming Guide
:INPut:CDR:LORDer[?]
Syntax
:INPut:CDR:LORDer 'identifier', <FIRSt|SECond>
:INPut:CDR[:STATe]? 'identifier'
Input
'identifier': 'M*.DataIn1' or 'M*.DataIn2'
Parameters
<FIRSt|SECond>: Set the loop order of the CDR.
FIRS|SEC
Description
This command determines the loop order of the CDR. Depending on this
setting, defining the behavior of the CDR is done by different parameters.
For first order only, transition density and bandwidth fully define its
characteristics while for the second order this definition needs at least one
more parameter i.e. peaking.
This SCPI is applicable for M8041A and M8051A.
Example
:INP:CDR:LORD 'M1.DataIn1', FIRS
:INPut:CDR:CTRL[?]
Syntax
:INPut:CDR:CTRL 'identifier', <MANual|SEQuence>
:INPut:CDR:CTRL? 'identifier'
Input
'identifier': 'M*.DataIn1' or 'M*.DataIn2'
Parameters
<MANual|SEQuence>: Set the CDR control method.
MAN|SEQ
Description
This command selects whether the CDR is controlled manually or by the
sequencer.
This SCPI is applicable for M8041A and M8051A.
Example
:INP:CDR:CTRL 'M1.DataIn1', SEQ
SCPI Command Reference