Section 3: Electrical installation
June 2017
Serial connections
Use serial cable (Daniel P/N 3-2500-401) to connect to a PC running Daniel MeterLink. The cable
is designed for RS-232 communications which is the serial Port A default configuration
Appendix A
plugs directly into the PC running Daniel MeterLink. The three wires on the other end of the
cable connect to the CPU Module. The RED wire goes to RX, the WHITE wire goes to TX, and the
BLACK wire goes to COM for the RS-485/RS-232
When Belden wire No. 9940 or equivalent is used, the maximum cable length for RS-232
communications at 9600 bps is 88.3 meters (250 ft.) and the maximum cable length for RS-485
communication at 57600 bps is 600 meters (1970 ft.).
Port A supports a special override mode which forces the port to use known communication
values (19200 baud, address 32, RS-232). Note that the protocol is auto-detected. This mode is
expected to be used during meter commissioning (to establish initial communication) and in
the event that the user cannot communicate with the meter (possibly due to an inadvertent
communication configuration change). Alternately, when using Daniel MeterLink™ with an
Ethernet port, use Ethernet cable (Daniel P/N

Table 3-4 Serial Port A parameters

Port/Communication
Port A (Standard)
RS-232
RS-485 Half Duplex
RS-485 Full Duplex
Ethernet
1.
Denotes auto-detected protocols.
If not using Ethernet, a full duplex serial connection is necessary for Daniel MeterLink to
communicate with a Daniel 3812 Liquid Ultrasonic Flow Meter.
56
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
field wiring diagram, Daniel Drawing DMC - 005558). The DB-9 end of the cable
Description
Typically used for general communications
with a flow computer, RTU (Modbus slave) and
radios.
Special override mode to force port configura-
tion to known settings.
Supports RTS/CTS handshaking with
software-configurable RTS on/off delay times.
Factory default is RS-232, Address 32, 19200
baud.
Preferred port for diagnostic communication
via Daniel MeterLink
10 Mbps/100 Mbps
(Table 3-4
for Port A wiring).
) to connect the PC.
1-360-01-596
Common features
Communications via Daniel MeterLink
using RS-232 or RS-485 Full Duplex
Software configurable Modbus Address
(1-247)
Auto-detects TCP/IP and ASCII or RTU
Protocol
-
ASCII Protocol:
Start Bits = 1, Data Bits=7
-
Parity: odd or even 1, Stop Bits =1
-
Baud Rates: 1200, 2400, 9600,
19200, 38400, 57600, 115000 bps
-
RTU Protocol:
Start Bits = 1, Data Bits=8
-
Parity: none, Stop Bits = 1
-
Baud Rates: 1200, 2400, 9600,
19200, 38400, 57600, 115000 bps
Modbus TCP/IP
CPU Module labeling and LED indicators
3-9000-765 Rev H
(see
1
1
1
1