Figure 3-7:
A. Cable ties
B. 3410 Series Ultrasonic Meter transducer port
C. Transducer assembly
WARNING!
CRUSHING HAZARD
During meter installation or removal, always place the unit on a stable platform or surface that
supports its assembled weight.
Failure to comply could allow the meter to roll, resulting in serious injury or equipment
damage.
3.4.1

Replace transducer cables

The meter body ports are identified by stamped or cast lettering adjacent to the
transducer port. Transducers for transmitter head 1 will be labeled 1-A1, 1-A2, etc. while
the transducer for transmitter head 2 will be labeled 2-A1, 2-A2, etc.
1.
2.
Maintenance and Troubleshooting manual
Daniel 3410 Series Ultrasonic Gas Flow Meter transducer cables and ports
Remove power to the meter.
Disconnect the transducer cable from the transformer retainer by removing the
retaining clips and pulling the cable plug straight out. Do not twist or rotate the
plug. (See
Figure
3-3.)
Meter repairs
29