Install the transducer holder

1.
2.
26
Figure 3-5:
Transducer holder length and set screw identification
A. Transducer holder set screws
B. Transducer holder length identification
Loosen the three Allen setscrews with a 1/16" hex driver securing the transducer
assembly and stalk, if installed. Carefully remove the transducer by pulling it from
the T-Slot transducer holder (or stalk if installed) without rotating.
Clean the holder with a dry cloth.
Ensure that the Daniel 3410 Series Ultrasonic Gas Flow Meter transducer port,
mount, and T-Slot transducer holder assembly are clean and free of debris.
Insert the transducer (parts are keyed and can only be assembled one way) into the
stalk or into the new transducer holder if no stalk is required. Do not use any
lubricant on the O-rings or contacts of the transducers.
NOTICE
Ensure that the transducers identified as belonging to end 1 are installed on end 1 of the
meter housing and those identified as belonging to end 2 are installed on end 2 of the
meter housing.