Seca 514 Administrator's Manual page 54

Body composition analyzer
Table of Contents

Advertisement

Problem
seca
No access to
patient
database of the seca 115 PC
software
USB memory stick connected
to touchscreen display but no
seca
access to the
patient
database
54 •
Cause
No
wireless network set up
seca 360°
between device and the PC on which
the
PC software is installed
seca 115
No Ethernet connection set up
between the device and the PC on
which the
PC software is
seca 115
installed
Device connected to a standalone PC
via Ethernet cable. The PC's network
card does not permit automatic
crossover
No USB memory stick connected to
the touchscreen display.
The PC on which the
seca 115
software is installed is not switched on
PC software not started
seca 115
No seca CLS network set up
seca CLS server of the
seca 115
software not started
Non-initialized USB memory stick being
used
PIN not entered or entered incorrectly
Unsuitable USB memory stick being
used
Interference from high-frequency emis-
sions from other devices (e.g. cell-
phones)
Administrator:
set up
seca 360°
"Setting up the seca 360° wireless
network" on page 40)
Administrator:
set up Ethernet connection (see "Integrate
device in an Ethernet network" on page 37)
Administrator:
use a crossover adapter (see "Technical
information" on page 56)
Connect USB memory stick to the
touchscreen display
PC
Switch on the PC and start the
PC software
Start the
seca 115
Administrator:
set up seca CLS network (see "Setting up
seca CLS network (via Ethernet only)" on
page 38)
PC
Administrator:
start seca CLS server manually
• Use the USB memory stick supplied
• Administrator: initialize USB memory stick
using the
seca 115
Use your user PIN or the PIN for the USB
memory stick.
• Use the USB memory stick supplied
• Use a FAT16 USB memory stick
Increase distance from high-frequency
devices
Remedy
wireless network (see
seca 115
PC software
PC software